Chapter 2 • CUSA
®
NXT System Setup and Components •
19
Initial Setup of Console with the Service Module
1.
Place Console in the recess on top of the Service Module.
2.
Plug the Service Module cord located in the module’s top rear
compartment into the power connector (
1
) on the rear of the Console.
3.
Plug one end of the detachable power cord into the Service Module’s
power connector (
2
) on the rear of the module.
4.
Connect the aspiration controller cable (
3
) located in the Service Module’s
top rear compartment to the port on the bottom of the Console (
4
) inside
the compartment
5.
Tighten both connector screws.
6.
Thread the locking thumbwheel located in the Service Module door into
the threaded hole on the bottom of the Console inside the compartment.
Tighten fully.
1
2
3
4