Chapter 6 • Disassembly and Sterilization of CUSA
®
NXT After Surgery •
81
Handling and Transporting the System
If you need to move/relocate the CUSA NXT System within the environment
of the Healthcare Facility, note the guidelines below.
The CUSA NXT System requires a minimum of one hour exposure at its
operating temperature range before you use it; make sure that the system is
located in the surgical room at least an hour before use.
Console with Service Module
Before moving the system:
•
Unplug the power cord from the wall receptacle and place the power cord
in the storage compartment at the rear of the Service Module.
•
Disconnect the handpiece.
•
Disconnect and remove the irrigation fluid.
•
Retract the IV pole.
•
Secure any loose objects. Make sure there are no loose objects placed on
top of the Console.
•
Disconnect the Footswitch and place the Footswitch in the storage
compartment at the rear of the Service Module.
When moving the system:
•
Push the system using the handle; don’t pull it.
•
Do not run whilst pushing the system.
•
Use a lift to move the system between floors of a building. Never use a
stairwell.
•
Do not attempt to lift the system.
To move the system up or down a ramp, use two or more people.
Console without Service Module
Before moving the Console:
•
Unplug the power cord from the wall receptacle. Disconnect the power
cord from the Console.
•
Disconnect the handpiece.
•
Disconnect and remove the irrigation fluid.
•
Retract the IV pole.
•
Secure any loose objects. Make sure there are no loose objects placed on
top of the Console.
•
Disconnect the footswitch.
When moving the Console (in the immediate environment):
•
Lift the Console using the side grips.
Summary of Contents for CUSA NXT
Page 3: ...This page is intentionally left blank...
Page 7: ...iv This page is intentionally left blank...
Page 39: ...32 Chapter 2 CUSA NXT System Setup and Components This page is intentionally left blank...
Page 73: ...66 Chapter 4 Preparing the CUSA NXT for Surgery This page is intentionally left blank...
Page 79: ...72 Chapter 5 Operating the CUSA NXT During Surgery This page is intentionally left blank...
Page 95: ...88 Chapter 7 Troubleshooting the CUSA NXT System This page is intentionally left blank...
Page 105: ...98 Appendix A Technical Specifications This page is intentionally left blank...
Page 113: ...106 Index This page is intentionally left blank...