6
• Chapter 1 • Patient and Operating Room Safety
Preparing the CUSA NXT System for Surgery
Caution
Pressing and releasing the
Power
pedal too quickly may result in a Handpiece
Tip Initialization failure. If this occurs, disconnect the handpiece, verify proper
tip assembly, and then reconnect the handpiece and try again.
Do not store anything in the storage area in the back of the Console while the
system is running.
Notice
After releasing the
Power
pedal, the pinch valve operates momentarily to block
aspiration and facilitate tissue release from the handpiece tip.
If you press the yellow symbol, the touch screen monitor will display error and
problem resolution information. For more information on responding to alarms,
see page 84.
The level of ultrasonic power displayed on the scale will only be directed to the
handpiece tip when the
power
Footswitch pedal is pressed fully. Pressing this
Footswitch pedal partially will give an actual power output below the setting
displayed with the exact value depending on the pedal position.
Note that the system only delivers irrigant to the handpiece tip when you press
the
Power
Footswitch pedal. When you press the
Fast Flush
Footswitch
pedal, the system ignores the
irrigation
scale setting and instead applies an
irrigation fast flush rate of about 50 ml/min to the handpiece.
Only the Console is shipped with a power cord. If a Service Module is
employed, this cord must be used for the Service Module instead. A lead comes
from the rear door of the Service Module and plugs into the back of the
Console.
Warning
Single Use devices are for single patient use only. Do not resterilize, reprocess
or re-use. Devices (s) is (are) intended to be used for one procedure only. If
reprocessed or re-used this may result in the infection of patient (or patient
specimen) through cross-contamination, as well as would incur the risk of
modifying the properties and performance of the device, and of increasing the
likelihood of complications and/or undesirable effects. Do not use devices which
have not been properly sterilized. Once used, devices must be disposed of in
accordance with hospital policies.
Using the incorrect wrench set to assemble/disassemble the handpiece may
damage the handpiece.
Failure to center the shroud may cause the handpiece to overheat.
Summary of Contents for CUSA NXT
Page 3: ...This page is intentionally left blank...
Page 7: ...iv This page is intentionally left blank...
Page 39: ...32 Chapter 2 CUSA NXT System Setup and Components This page is intentionally left blank...
Page 73: ...66 Chapter 4 Preparing the CUSA NXT for Surgery This page is intentionally left blank...
Page 79: ...72 Chapter 5 Operating the CUSA NXT During Surgery This page is intentionally left blank...
Page 95: ...88 Chapter 7 Troubleshooting the CUSA NXT System This page is intentionally left blank...
Page 105: ...98 Appendix A Technical Specifications This page is intentionally left blank...
Page 113: ...106 Index This page is intentionally left blank...