84
• Chapter 7 • Troubleshooting the CUSA
®
NXT System
Responding to Alarms
All alarms on the CUSA NXT are low-priority, technical alarms; operator
awareness is required.
Symbol
Alarm
Causes
Recommendations
Aspiration
Tubing not attached to suction
canister.
Attach the tubing to the suction
canister.
Power
Improper tip assembly.
• Tighten the tip.
• Reconnect the handpiece.
Bad tip.
• Replace the tip.
• Reconnect the handpiece.
Handpiece
Handpiece not connected.
Connect the handpiece to the
appropriate port:
• Red 24 kHz
• Green 35 kHz
• Black 24 kHz
Microsurgical
Two or more handpieces are
connected.
Disconnect extra handpieces.
Footswitch
Footswitch not connected.
Connect the Footswitch to the
rear of Console.
More than one Footswitch pedal
is pressed.
Release one Footswitch pedal.
Footswitch pedal is pressed
while connecting components to
the Console.
Release the Footswitch pedal.
General System
General system fault.
Restart the system. If the alarm
persists, contact technical
support.
Notice
If you press the yellow symbol, the touch screen monitor will display error
and problem resolution information.
Summary of Contents for CUSA NXT
Page 3: ...This page is intentionally left blank...
Page 7: ...iv This page is intentionally left blank...
Page 39: ...32 Chapter 2 CUSA NXT System Setup and Components This page is intentionally left blank...
Page 73: ...66 Chapter 4 Preparing the CUSA NXT for Surgery This page is intentionally left blank...
Page 79: ...72 Chapter 5 Operating the CUSA NXT During Surgery This page is intentionally left blank...
Page 95: ...88 Chapter 7 Troubleshooting the CUSA NXT System This page is intentionally left blank...
Page 105: ...98 Appendix A Technical Specifications This page is intentionally left blank...
Page 113: ...106 Index This page is intentionally left blank...