• Appendix A • Technical Specifications
Manufacturer’s Declaration Table
Notice
The CUSA NXT System should not be used adjacent to or stacked with
equipment other than the equipment specified in the CUSA NXT System
Operator’s Manual. If adjacent or stacked use is necessary, the CUSA NXT
System should be observed to verify normal operation in the configuration in
which it will be used. The use of accessories, other than the accessories
specified in the CUSA NXT System Operator’s Manual may result in increased
emissions or decreased immunity of the CUSA NXT System.
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The CUSA NXT System is intended for use in the electromagnetic environment specified below. The customer or the
user of the CUSA NXT System should assure that it is being used in such an environment.
Emissions Test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
The CUSA NXT System uses RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions CISPR 11
Class A
The CUSA NXT System is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions IEC 61000-3-2
Class A
Voltage fluctuations/flicker emissions IEC
61000-3-3
Complies with
Standard
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The CUSA NXT System is intended for use in the electromagnetic environment specified below. The customer or the
user of the CUSA NXT System should assure that it is being used in such an environment.
Immunity Test
IEC 60601 test level
Compliance
Electromagnetic environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
Complies with
Standard
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
transient/burst IEC
61000-4-4
±2 kV for power lines
±1 kV for input/output lines
Complies with
Standard
Mains power quality should be that of a typical
commercial or hospital environment.
Surge IEC
61000-4-5
±1 kV differential mode
±2 kV common mode
Complies with
Standard
Mains power quality should be that of a typical
commercial or hospital environment.
Summary of Contents for CUSA NXT
Page 3: ...This page is intentionally left blank...
Page 7: ...iv This page is intentionally left blank...
Page 39: ...32 Chapter 2 CUSA NXT System Setup and Components This page is intentionally left blank...
Page 73: ...66 Chapter 4 Preparing the CUSA NXT for Surgery This page is intentionally left blank...
Page 79: ...72 Chapter 5 Operating the CUSA NXT During Surgery This page is intentionally left blank...
Page 95: ...88 Chapter 7 Troubleshooting the CUSA NXT System This page is intentionally left blank...
Page 105: ...98 Appendix A Technical Specifications This page is intentionally left blank...
Page 113: ...106 Index This page is intentionally left blank...