Chapter 3 • Using the Console, Service Module, and Handpieces •
41
Adjusting the Ultrasonic Control Scales
Use the
power
,
irrigation
, and
aspiration
scales to control the major
ultrasonic settings during a case.
Adjusting the Power Level
Use the
power
scale (orange) to specify the ultrasonic power output (tip
amplitude). The power settings range between 10% and 100%. The default
power setting is 10%.
Note that raising the power setting will increase
the vibration of the handpiece tip and thus
increase tissue fragmentation rate. To adjust the
power by 5% increments, use the arrows to the
right of the scale. To adjust the power by 10%
increments, press the segment bars on the scale.
Notice
The level of ultrasonic power displayed on the scale
will only be directed to the handpiece tip when the
power
Footswitch pedal is pressed fully. Pressing
this Footswitch pedal partially will give an actual
power output below the setting displayed with the
exact value depending on the pedal position.
Summary of Contents for CUSA NXT
Page 3: ...This page is intentionally left blank...
Page 7: ...iv This page is intentionally left blank...
Page 39: ...32 Chapter 2 CUSA NXT System Setup and Components This page is intentionally left blank...
Page 73: ...66 Chapter 4 Preparing the CUSA NXT for Surgery This page is intentionally left blank...
Page 79: ...72 Chapter 5 Operating the CUSA NXT During Surgery This page is intentionally left blank...
Page 95: ...88 Chapter 7 Troubleshooting the CUSA NXT System This page is intentionally left blank...
Page 105: ...98 Appendix A Technical Specifications This page is intentionally left blank...
Page 113: ...106 Index This page is intentionally left blank...