38
• Chapter 3 • Using the Console, Service Module, and Handpieces
Activating the Handpiece
To activate the handpiece, press the
Power
pedal for at least 2 seconds.
Pressing the Fast Flush Pedal
Pressing this pedal will:
•
Switch on irrigation at a rate of 50 ml/min. Note that the irrigation setting
on the front panel display has no influence upon the irrigation rate when
the
Fast Flush
pedal is pressed independently.
•
Stop aspiration (until the pedal is released).
Using the Touch Screen Monitor
The CUSA NXT Console provides a touch screen monitor for viewing and
controlling the power, irrigation, and aspiration levels during an ultrasonic
case.
Caution
Pressing and releasing the
Power
pedal too quickly may result in a Handpiece
Tip Initialization failure. If this occurs, disconnect the handpiece, verify proper
tip assembly, and then reconnect the handpiece and try again.
1
2
3
4