The Service Module includes a spare CUSA NXT contamination guard. For
instructions on setting up the Service Module, see page 59.
Collecting Waste
A receptacle for collecting waste is required between the CUSA NXT
Console and the vacuum source. RECEPTAL
®
or Bemis
®
Hi-Flow™ 2000
ml canisters are suggested.
Using Other Suction Source Types
If the CUSA NXT Console will not be used with the Service Module, then the
Console will require another suitable source of suction and waste collection.
Wall or ceiling mounted vacuum points in the operating room or a
free-standing surgical suction unit should operate at a constant vacuum
pressure between 560 mmHg and 620 mmHg with a free air flow capacity of
more than 42 L/min.
Warning
TO AVOID RISK OF ELECTRIC SHOCK, THIS EQUIPMENT MUST ONLY BE
CONNECTED TO A SUPPLY MAINS WITH PROTECTIVE EARTH.