Chapter 4 • Preparing the CUSA
®
NXT for Surgery •
63
3.
On the left side of the Console, connect the aspiration filter to the
designated port (
2
).
Position the Footswitch
Connect the footswitch, see page 36. Position the footswitch where it is easily
accessible to the surgeon.
Priming the Handpieces
Automatic priming is a feature on the CUSA NXT System. Manual methods
for priming used with previous versions of Selector may still be performed,
however, you must at least start and then stop the initial prime cycle in order
to proceed with system operation. If any alarm conditions exist upon initial
startup (e.g. handpiece or footswitch not connected), the
prime
screen will
not appear until these alarms are resolved.
Notice
Do not remove the aspiration filter from the designated port prior to powering off
the system.
1
2