Chapter 6 • Disassembly and Sterilization of CUSA
®
NXT After Surgery •
75
Cleaning the Handpieces After Surgery
Disassembling the Handpieces
After surgery and prior to sterilization, disassemble and clean the CUSA
®
Selector
®
handpiece:
1.
In a sink or bowl, prepare a cleaning solution containing warm water with
detergent. This is necessary for cleaning the handpiece parts. Consult
hospital policy for approved cleaning materials.
2.
Remove and dispose of the clear flue after use, in compliance with the
hospital policy for disposal of contaminated waste.
3.
Remove the black shroud and place in the cleaning solution.
4.
Lay the handpiece in the appropriate wrench holder with the wrench flat
secured in the holder cut-out.
5.
Using the appropriately sized end of the wrench, loosen the tip by turning
it counterclockwise. Unscrew the tip and place it in the cleaning solution.
If the tip is single-use only or shows evidence of damage, discard the tip.