Chapter 1 • Patient and Operating Room Safety •
5
Using the Console, Service Module, and Handpieces
Only use the CUSA Selector wrench kit to tighten tips and extensions on the
transducer of the handpiece. Always secure the handle of the handpiece in the
holder provided. Never dismantle or assemble the handpiece without placing
the handle section of the handpiece in the holder. Failure to comply with this
instruction could result in permanent damage to the transducer and will
invalidate the warranty.
Do not attempt to clear a blocked handpiece by passing the stylet down the rear
inlet of the handpiece. This may result in damage to the handpiece. Always
pass the stylet from the tip end.
Do not use a
handpiece with a broken tip. Inspect the tip for damage prior to
use. If the tip is damaged it may be replaced by the operator.
Do not use dry air sterilization methods involving temperatures in excess of
139 °C (282 °F); this may damage the handpiece.
Do not power up the CUSA Selector handpiece without fully priming the
handpiece with irrigant (i.e. irrigant must flow from the tip of the handpiece).
Failure to prime the handpiece could result in permanent damage to the
handpiece (internal and/or external components) and may induce heat
exchange with the patient or user, and cause tissue burns.
Notice
In the event that you need to return a device to Integra, to prevent any damage,
use the appropriate packaging material.
To prevent fluid flowing into the vacuum line, only use a canister that has a non
return valve.
Caution
Warning
When you connect the handpiece to the Console, the handpiece becomes a
functional surgical device.
Ignoring alarms on the CUSA NXT System while continuing to use the system
may result in injury to the patient and/or surgical personnel, or equipment
damage.
The minimum aspiration level setting is 15%. This ensures proper handpiece
cooling and performance. Insufficient aspiration can cause premature
handpiece or tip failures and may induce heat exchange with the patient or user,
and cause tissue burns.
Summary of Contents for CUSA NXT
Page 3: ...This page is intentionally left blank...
Page 7: ...iv This page is intentionally left blank...
Page 39: ...32 Chapter 2 CUSA NXT System Setup and Components This page is intentionally left blank...
Page 73: ...66 Chapter 4 Preparing the CUSA NXT for Surgery This page is intentionally left blank...
Page 79: ...72 Chapter 5 Operating the CUSA NXT During Surgery This page is intentionally left blank...
Page 95: ...88 Chapter 7 Troubleshooting the CUSA NXT System This page is intentionally left blank...
Page 105: ...98 Appendix A Technical Specifications This page is intentionally left blank...
Page 113: ...106 Index This page is intentionally left blank...