Chapter 3 • Using the Console, Service Module, and Handpieces •
39
This touch screen provides the following tools:
For instructions on maintaining the touch screen, see page 80.
Status/Alarm Symbols
The CUSA NXT System activates an alarm to indicate a technical problem
with the system. All alarms on the CUSA NXT are technical, low-priority
alarms, for example, mechanical or equipment related. There are no
physiological alarms on the CUSA NXT System.
The five symbols that appear on top of the touch screen provide status
information on the major ultrasonic components.
Tools
Description
1
Status / Alarm Symbols
Displays colored symbols to indicate the status of the major ultrasonic
components. For specific information, see the page 39.
2
Ultrasonic Control Scales
Controls the power, irrigation, and aspiration settings during an ultrasonic case
(see page 41).
3
Information Panel
Displays general information regarding the accumulated ultrasonic time and the
current vacuum levels (see page 43).
4
Menu Buttons
Displays menus for priming the handpieces, customizing user settings, and
viewing online help (see page 44).
Symbol
Normal
Status
Alarm Indicator (When Yellow)
Alarm
Classification
Alarm
Priority
Aspiration
Signifies an error in the aspiration
process.
Technical
Low
Power
Signifies an error in the ultrasonic
generator.
Technical
Low
Footswitch
Signifies a fault in the Footswitch or
associated wiring.
Technical
Low
Handpiece
Signifies that a handpiece is not
connected.
Technical
Low
General
Signifies an error in the electronics or
power supplies.
Technical
Low
Summary of Contents for CUSA NXT
Page 3: ...This page is intentionally left blank...
Page 7: ...iv This page is intentionally left blank...
Page 39: ...32 Chapter 2 CUSA NXT System Setup and Components This page is intentionally left blank...
Page 73: ...66 Chapter 4 Preparing the CUSA NXT for Surgery This page is intentionally left blank...
Page 79: ...72 Chapter 5 Operating the CUSA NXT During Surgery This page is intentionally left blank...
Page 95: ...88 Chapter 7 Troubleshooting the CUSA NXT System This page is intentionally left blank...
Page 105: ...98 Appendix A Technical Specifications This page is intentionally left blank...
Page 113: ...106 Index This page is intentionally left blank...