92
• Appendix A • Technical Specifications
voltage setting switch is located inside the rear compartment under the
access panel. The access panel displays two stamps, 115 V and 230 V, at
opposite corners of the access panel. The position of the access panel
determines what voltage the Service Module is set at. The maximum current
drawn is 2.5 amps at 240 V and 5 amps at 100 V. The removable power cord
supplied with the system provides the power.
Correct grounding of the CUSA NXT Console is essential for safe operation.
The Console is grounded via the power cord; verify that this cord is
connected to a properly grounded power receptacle of appropriate rating.
Only use the power cord supplied with the system since these cords vary
according to destination. An external ground point is also located on the rear
panel (see page 35).
Warning
TO AVOID RISK OF ELECTRIC SHOCK, THIS EQUIPMENT MUST ONLY BE
CONNECTED TO A SUPPLY MAINS WITH PROTECTIVE EARTH.