19 Turn on Micro GC Fusion. The instrument is ready to perform analysis.
CAUTION
If the flow rate through the device exceeds 100 mL/min, the output pressure
from the device to Micro GC Fusion may exceed the acceptable pressure
range, causing instrument damage.
WARNING
Safely vent carrier and sample streams - potentially toxic, noxious, asphyxiant
(oxygen displacing) or flammable gases - outside Micro GC Fusion and away
from the operating area. Vent toxic gases to a fume hood, chemical trap, or
reaction medium.
6.15.2 4-Module Micro GC Fusion Chassis Installation - Side
Install
The pressure reducer and gas-liquid separator includes the following items. These
items are needed to install the device on the side of a 4-module Micro GC Fusion
chassis.
PN
Description
Quantity
059-0696
5/16 in. female nut
1
059-0697
1/16 in. ferrule set
1
059-0336
1/16 in. Valco male nut
1
059-0335
1/16 in. ferrule
1
952-229-G1
sample tube
1
pressure reducer and gas-
liquid separator assembly
1
There are three tubes in kit (PN 952-229-G1) and the longest length tube is
used for this installation.
Tools required:
• 3/8 in. open-ended wrench
• 5/16 in. open-ended wrench
• 1/4 in. open-ended wrench