10 | Routine Operation
INFICON
262 / 319
074-594-P1H Micro GC Fusion Operating Man-
ual
11 Remove the retaining screws:
ð
For modules C/D, remove the retaining screw holding the JP-100 in place
with the 2.5 mm ball point hex driver.
ð
For modules A/B, remove the retaining screw holding the JP-100 in place with
the 2.5 mm ball point hex wrench.
12 Slide the JP-100 up and out of the instrument.
13 See GC Module Removal Procedure for 4-Module System [
the module removal process.