9.10.10.8.2 Multi-Level Calibration
Create several sequences corresponding to the calibration level using the method
described in Single Level Calibration [
228].
Example - An auto-calibration sequence for level 3 would be named Level 3
Calibration and have a 3 typed in the Calibration Runs Level field.
9.10.10.8.3 Multi-Level Calibration using a Valco Stream Selector
1 Create a sequence called Auto-Calibration. See Add Sequence [
170].
2 Add a Valco Stream Selector switching event and switch to the first calibration
gas.
3 Click the icon and select Calibration Runs.
4 Type the number of step repetitions and the method. If desired, add a delay
between each repetition.
5 Type the number of runs to use while averaging the area.
6 Type 1 for a level 1 calibration.
7 Click the icon and select Add Valve.
8 Use the valve menu to select the level 2 calibration gas.
9 Select Calibration Runs.