6.2.3 Making Swagelok Connections
A 1/16 in. sample inlet Swagelok connector is located on the Micro GC Fusion
faceplate. A 1/8 in. carrier gas Swagelok connector is located on the back of Micro GC
Fusion. The vent ports on the back of Micro GC also use 1/8 in. Swagelok connectors.
This section reviews how to construct Swagelok connections and connect tubing to a
fitting, such as the sample inlet(s), carrier gas inlet(s), vents, or the gas supply tank.
Materials Needed:
• 1/8 in. OD pre-cleaned copper or stainless steel tubing
• 1/8 in. Swagelok nuts, and front and back ferrules
• Two 7/16 in. wrenches for 1/8 in. tubing
• Two 5/16 in. wrenches for 1/16 in. tubing
Procedure
1 Attach a 1/8 in. Swagelok nut, back ferrule and front ferrule to the tubing.
Female fitting (1/8 in.)
Front ferrule (1/8 in.)
Back ferrule (1/8 in.)
Swagelok nut (1/8 in.)
Tubing(1/8 in.)
2 Clamp a female fitting in a bench vise.
3 Push the tubing into the female fitting.