INFICON
Installation | 6
074-594-P1H Micro GC Fusion Operating Manual
79 / 319
12 Connect the sample output tubing from the gas-liquid separator to the Micro GC
Fusion sample inlet and tighten using a 5/16 in. wrench.
13 Tighten the Valco fitting on the gas-liquid separator using a 1/4 in. wrench.
14 Use additional tubing to connect the gas-liquid separator sample vent to a fume
hood or other safe location.
15 Reconnect all electrical cables, including the power cable and Ethernet cable.
16 Reconnect the carrier gas and venting connections.
17 Turn on Micro GC Fusion. Connect a sample to the sample inlet on gas-liquid
separator to perform an analysis.
WARNING
Safely vent carrier and sample streams - potentially toxic, noxious, asphyxiant
(oxygen displacing) or flammable gases - outside Micro GC Fusion and away
from the operating area. Vent toxic gases to a fume hood, chemical trap, or
reaction medium.