Troubleshooting | 11
074-594-P1H Micro GC Fusion Operating Manual
285 / 319
Symptom
Possible sources of error
Remedy
There is an internal leak.
• Contact INFICON. See Service and
There is a malfunction in the
column or injector.
• Check the flow rate of the columns by
attaching a flow meter to the analytical and
reference vents on the back of Micro GC
Fusion
• If there is no flow from either or both vents,
contact INFICON. See Service and
Technical Support [
There is a malfunction in the
detector.
• Contact INFICON. See Service and
The chromatogram is
displaying a flat line with
no peaks
The Inject (ms) time method
parameter is set to 0 ms.
• Change the Inject (ms) time to a value other
than 0 ms.
The Sample Pump Time method
parameter is set to 0 seconds.
• Ensure that the Sample Pump Time is set for
at least 15 seconds.
There is a malfunction in the
column or injector.
• Check the flow rate of the columns by
attaching a flow meter to the analytical and
reference vents on the back of Micro GC
Fusion
• If there is no flow from either or both vents,
contact INFICON. See Service and
Technical Support [
There is a malfunction in the
sample pump or the sample
pump does not turn on.
• Listen for the noise from the pump when
taking a sample. If there is no noise, contact
INFICON.
There is a malfunction in the
detector.
• Contact INFICON. See Service and
A peak is missing
The Backflush Time method
parameter is set too early (if
applicable).
• Increase the Backflush Time until all peaks
of interest appear on the chromatogram.
A peak is saturated
The Inject (ms) time method
parameter is set too long.
• Shorten the Inject (ms) time.
The Injector Temperature
method parameter is set too
high.
• Lower the Injector Temperature.
Chromatograms are not
appearing/no data is
showing
The web browser requires a
restart or refresh.
• Restart or refresh the web browser.