ZERO
Edge-controlled input
Change from LOW to HIGH: Activate ZERO.
Change from HIGH to LOW: Deactivate ZERO
START/STOP
Edge-controlled input
Change from LOW to HIGH: activate START
Change from HIGH to LOW: activate STOP
START
Edge-controlled input
Change from LOW to HIGH: carry out START
STOP
Edge-controlled input
Change from LOW to HIGH: activate STOP
If this input is HIGH for a longer period than it is set in "Ventilation Delay", ventilate
additionally.
VENT
Edge-controlled input
Change from LOW to HIGH: Flooding
GAS BALLAST
Edge-controlled input
Change from LOW to HIGH: Activate Gas Ballast/Purge.
Change from HIGH to LOW: Deactivate Gas Ballast/Purge.
CLEAR
Edge-controlled input
Change from LOW to HIGH: Acknowledge error message or abortion of a calibration.
CAL
Edge-controlled input
Change from LOW to HIGH:
In case the device is in Standby mode: Start internal automatic calibration. In case
the machine is in Measurement mode: Start manual calibration. (Precondition: exter-
nal test leak has to be open and leak rate signal is constant)
Change from HIGH to LOW:
External calibration: approve that external test leak is closed and leak rate signal is
stable.
Summary of Contents for 550-300A
Page 6: ...1 4 Content jinb80e1 h Operating Manual 1501 ...
Page 10: ...1 8 Operating instructions jinb80e1 h Operating Manual 1501 ...
Page 18: ...2 8 Important safety instructions jinb80e1 h Operating Manual 1501 ...
Page 26: ...3 8 Description of equipment jinb80e1 h Operating Manual 1501 ...
Page 111: ...Technical Data 9 23 jinb80e1 h Operating Manual 1501 9 7 Commander mode Fig 9 14 Flow Chart ...
Page 112: ...9 24 Technical Data jinb80e1 h Operating Manual 1501 9 8 CE Certificate Fig 9 15 ...
Page 122: ...10 10 Error Messages and Warnings jinb80e1 h Operating Manual 1501 ...
Page 124: ...11 2 Ordering Information jinb80e1 h Operating Manual 1501 ...