Reference measurement
In this menu, the reference measurement can be started.
Main menu
o
Settings
o
Vacuum settings
o
Commander Functions
o
In the menu item „Commander Functions“, all parameters of the commander function
can be configured.
o
Commander Timing
The individual process times can be set here.
The adjustment range is from 0.1 to 999.9 seconds
t_A evacuation time
Maximum time to achieve the set sample pressure "p_B evacuation pressure"
Default setting: 30 seconds
t_B zero delay
Time delay until ZEROING (if zero "enabled") or time in which I•ZERO is carried out
(if I·ZERO has been activated);
Default setting: 10 seconds
t_C filling time
Maximum time to achieve the filling pressure "p_C Filling pressure" in the unit under
test
Default setting: 30 seconds
t_D depressurising time
Maximum time to achieve the set depressurising pressure "p_B evacuation pres-
sure"
; factory setting: 30 seconds
t_E flooding time
Maximum time to achieve the set flooding pressure p_A gross leak test
; factory setting: 10 seconds
t_F readiness for measuring
Maximum time to achieve the Measuring mode
; factory setting: 10 seconds
t_G measuring time
Time, after which the leak rate signal has been balanced.
; factory setting: 10 seconds
Summary of Contents for 550-300A
Page 6: ...1 4 Content jinb80e1 h Operating Manual 1501 ...
Page 10: ...1 8 Operating instructions jinb80e1 h Operating Manual 1501 ...
Page 18: ...2 8 Important safety instructions jinb80e1 h Operating Manual 1501 ...
Page 26: ...3 8 Description of equipment jinb80e1 h Operating Manual 1501 ...
Page 111: ...Technical Data 9 23 jinb80e1 h Operating Manual 1501 9 7 Commander mode Fig 9 14 Flow Chart ...
Page 112: ...9 24 Technical Data jinb80e1 h Operating Manual 1501 9 8 CE Certificate Fig 9 15 ...
Page 122: ...10 10 Error Messages and Warnings jinb80e1 h Operating Manual 1501 ...
Page 124: ...11 2 Ordering Information jinb80e1 h Operating Manual 1501 ...