Replace Vent Filter
The flood filter is checked and replaced in the course of the 5000-hours mainte-
nance. In case of extreme operating conditions shorter maintenance times could be
reasonable.
Required Tools
• Allen key SW 8
Required material
• Replacement filter (2 pcs.) Order No.: 200 000 683
Access to the flood filter is only possible when the cover of the Modul1000 has been
removed. Follow the instructions in
7.5.1
Opening of the Device
.
Notice:
Please observe the safety instructions in this chapter.
• To dismantle the vent filter, disconnect the hose connections of the filter from the
quick-connecting couplings. By pressing on the outside ring on the quick
connecting couplings, the hose can be pulled off. Please find the mounting
positions in the following
Fig. 7-8
.
Fig. 7-8 Flood filter replacement in Modul1000
Item Description
Item Description
1
Vent filter
2
Hose routing
a
1
a
Summary of Contents for 550-300A
Page 6: ...1 4 Content jinb80e1 h Operating Manual 1501 ...
Page 10: ...1 8 Operating instructions jinb80e1 h Operating Manual 1501 ...
Page 18: ...2 8 Important safety instructions jinb80e1 h Operating Manual 1501 ...
Page 26: ...3 8 Description of equipment jinb80e1 h Operating Manual 1501 ...
Page 111: ...Technical Data 9 23 jinb80e1 h Operating Manual 1501 9 7 Commander mode Fig 9 14 Flow Chart ...
Page 112: ...9 24 Technical Data jinb80e1 h Operating Manual 1501 9 8 CE Certificate Fig 9 15 ...
Page 122: ...10 10 Error Messages and Warnings jinb80e1 h Operating Manual 1501 ...
Page 124: ...11 2 Ordering Information jinb80e1 h Operating Manual 1501 ...