Maintenance tasks
7-13
jinb
80
e1
-h
Op
er
at
in
g
M
anua
l
(1
50
1)
7.8
Replacing of parameter memory (I•STICK)
The customer's application parameters are also stored in I•STICK If a backup unit is
required, the application parameters can easily be transferred by connecting the
I•STICK to the backup unit.
Required Tools
• Allen key SW 8
• Screwdriver size 1
Required material
• I•STICK
Access to the I•STICK is only possible when the cover of the Modul1000 has been
removed. Follow the instructions in
7.5.1
Opening of the Device
.
Notice:
Please observe the safety instructions in this chapter.
• To replace the I•STICK, release the two screws with which the I•STICK is
screwed on the connector. See the
Fig. 7-7
below.
After the screws have been released, pull the I•STICK out of the socket and insert
the I•STICK of the defect device.
The installation is performed in reverse order.
Fig. 7-7 I•STICK replacement
Item Description
Item Description
1
I•STICK
2
Fixing screws
1
2
Summary of Contents for 550-300A
Page 6: ...1 4 Content jinb80e1 h Operating Manual 1501 ...
Page 10: ...1 8 Operating instructions jinb80e1 h Operating Manual 1501 ...
Page 18: ...2 8 Important safety instructions jinb80e1 h Operating Manual 1501 ...
Page 26: ...3 8 Description of equipment jinb80e1 h Operating Manual 1501 ...
Page 111: ...Technical Data 9 23 jinb80e1 h Operating Manual 1501 9 7 Commander mode Fig 9 14 Flow Chart ...
Page 112: ...9 24 Technical Data jinb80e1 h Operating Manual 1501 9 8 CE Certificate Fig 9 15 ...
Page 122: ...10 10 Error Messages and Warnings jinb80e1 h Operating Manual 1501 ...
Page 124: ...11 2 Ordering Information jinb80e1 h Operating Manual 1501 ...