Technical Data
9-5
jinb
80
e1
-h
Op
er
at
in
g
M
anua
l
(1
50
1)
CAL INT
Edge-controlled input
Change from LOW to HIGH: start internal automatic calibration.
CAL EXT
Edge-controlled input
Change from LOW to HIGH: Start external manual calibration.
CYCLE
(inverted START / STOP input)
Status-controlled input
Change from LOW to HIGH: In the Standby status, START will be executed and in
the Measurement status, STOP will be executed.
GAS BALLAST ON
Edge-controlled input
Change from LOW to HIGH: The gas ballast valve is opened.
GAS BALLAST OFF
Edge-controlled input
Change from LOW to HIGH: The gas ballast valve is closed.
ZERO ON
Edge-controlled input
Change from LOW to HIGH: Activate ZERO
ZERO is switched off, when the devices switches over to Standby.
SNIFF
Status-controlled input
Change from LOW to HIGH in Standby mode:
Switching over to the SNIFF mode.
Change from HIGH to LOW in Standby mode:
Switching over to the set operating mode.
Summary of Contents for 550-300A
Page 6: ...1 4 Content jinb80e1 h Operating Manual 1501 ...
Page 10: ...1 8 Operating instructions jinb80e1 h Operating Manual 1501 ...
Page 18: ...2 8 Important safety instructions jinb80e1 h Operating Manual 1501 ...
Page 26: ...3 8 Description of equipment jinb80e1 h Operating Manual 1501 ...
Page 111: ...Technical Data 9 23 jinb80e1 h Operating Manual 1501 9 7 Commander mode Fig 9 14 Flow Chart ...
Page 112: ...9 24 Technical Data jinb80e1 h Operating Manual 1501 9 8 CE Certificate Fig 9 15 ...
Page 122: ...10 10 Error Messages and Warnings jinb80e1 h Operating Manual 1501 ...
Page 124: ...11 2 Ordering Information jinb80e1 h Operating Manual 1501 ...