Error Messages and Warnings
10-3
jinb
80
e1
-h
Op
er
at
in
g
M
anua
l
(1
50
1)
E34
24 V voltage at MSV board is
too low!
Signal MVPZN on the MSV board is active. 24 V voltage on the MSV board is too
low, U <18.3 V.
Fuse F1 on the MSV board has burned through.
Reference voltage UREF on the MSV board XT7/1 is too high, U>5 V.
DC/DC transformer of MSV board is defective
24 V current supply from the main power supply unit is excessively strained or
defective.
E35
Anode - cathode voltage is too
high!
Anode-cathode voltage is higher than U > 130 V.
MSV is faulty.
E36
Anode - cathode voltage is too
low!
Anode-cathode voltage is lower than U < 30 V.
Fuse F4 on the MSV board is defective.
MSV is faulty.
E37
Reference size of suppressor
voltage is too great.
Signal MFSZH on MSV board is active. Suppressor signal control input is too high.
Suppressor voltage is short circuited (cable, ion accumulator)
MSV is faulty.
E38
Suppressor potential is too high.
Suppressor potential is greater than 363V.
MSV is defective.
E39
Suppressor potential is too low.
Suppressor potential is lower than U < 297 V.
Short circuit in suppressor line.
MSV is faulty.
High-impedance short circuit in the ion collector.
E40
The anode potential exceeds
the target value by more than
10%.
The actual anode potential exceeds its nominal value by 10%. The nominal value
can be displayed in the service menu (in "Info")
MSV is faulty.
MC 68 is defective
E41
The anode potential falls below
the target value by more than
10%.
The actual anode potential has dropped below its nominal value by 10%. The
nominal value can be displayed in the service menu (in "Info")
Brief pressure increase in the mass spectrometer.
MSV is faulty.
MC 68 is defective
E42
Target value of the anode
potential is too high.
Signal MFAZH on MSV board is active. Reference size of the anode potential is
too great.
Brief pressure increase in the mass spectrometer.
Valve contamination causes briefly a high mass spectrometer pressure.
Anode voltage has been short circuited.
Nominal value of the anode voltage is too high. Anode voltage is limited to about
1,200 V.
E43
Cathode current is too high.
Signal MPKZH on MSV board is active. Cathode current is too high, I >3.6 A.
MSV board is defective.
No.
Displayed Message
Description and possible solutions
Summary of Contents for 550-300A
Page 6: ...1 4 Content jinb80e1 h Operating Manual 1501 ...
Page 10: ...1 8 Operating instructions jinb80e1 h Operating Manual 1501 ...
Page 18: ...2 8 Important safety instructions jinb80e1 h Operating Manual 1501 ...
Page 26: ...3 8 Description of equipment jinb80e1 h Operating Manual 1501 ...
Page 111: ...Technical Data 9 23 jinb80e1 h Operating Manual 1501 9 7 Commander mode Fig 9 14 Flow Chart ...
Page 112: ...9 24 Technical Data jinb80e1 h Operating Manual 1501 9 8 CE Certificate Fig 9 15 ...
Page 122: ...10 10 Error Messages and Warnings jinb80e1 h Operating Manual 1501 ...
Page 124: ...11 2 Ordering Information jinb80e1 h Operating Manual 1501 ...