Installation
4.1
Mechanical Installation
The Modul1000 can be mounted on and below desktop workplaces. The leak test
device may only be operated on horizontal surfaces. During the installation, make
sure that the Modul1000 gets enough fresh air, especially the air inlet and outlet
openings on the front and the back side have to be free. The maximum permissible
ambient temperature of the Modul1000 may not be exceeded during operation!
If the leak detector Modul1000 is used for an application that can cause strong vibra-
tions or impacts, then we recommend connection via flexible connections and to pref-
erably avoid direct connections. Please also use also flexible connections for the test
chambers, angle valves and the like if wear from impact or vibration is to be
expected.
The following figure illustrates the recommended impact or vibration direction for the
use of angle valves, test chambers or similar accessories which can cause unsuita-
ble strain for Modul1000:
Warning
Back injury through heavy loads.
Two people should carry the Modul1000 or use a lifting device.
Summary of Contents for 550-300A
Page 6: ...1 4 Content jinb80e1 h Operating Manual 1501 ...
Page 10: ...1 8 Operating instructions jinb80e1 h Operating Manual 1501 ...
Page 18: ...2 8 Important safety instructions jinb80e1 h Operating Manual 1501 ...
Page 26: ...3 8 Description of equipment jinb80e1 h Operating Manual 1501 ...
Page 111: ...Technical Data 9 23 jinb80e1 h Operating Manual 1501 9 7 Commander mode Fig 9 14 Flow Chart ...
Page 112: ...9 24 Technical Data jinb80e1 h Operating Manual 1501 9 8 CE Certificate Fig 9 15 ...
Page 122: ...10 10 Error Messages and Warnings jinb80e1 h Operating Manual 1501 ...
Page 124: ...11 2 Ordering Information jinb80e1 h Operating Manual 1501 ...