Opening of the Device
Required Tools
Allen wrench 8 mm
Open the Modul1000 as follows:
1
Switch off the line switch and remove the power cable.
2
Remove the clamp ring for quick release.
3
Use the Allen wrench SW 8 mm and open the locks (ROTO-LOCK) on both sides of
the device cover. Refer to
Fig. 7-2
.
1
Turn the Roto-Locks until they reach their end positions „OPEN“.
2
Carefully lift the cover.
3
The cover is reassembled in reverse order. When attaching the cover ensure that no
electrical wires reach between the cover and the chassis.
4
To secure the cover, turn the Roto-Locks until they reach their end positions
„CLOSE“.
STOP
Danger
Before removing the cover of the device you must disconnect the leak test device
from the mains cord.
Fig. 7-2 Open the cover