Technical Data
9-1
jinb
80
e1
-h
Op
er
at
in
g
M
anua
l
(1
50
1)
9
Technical Data
9.1
Technical data of components
9.1.1
Power supply
9.1.2
Weight / dimensions
9.1.3
Characteristics
Mains voltage and frequencies
100 V…240 V, 50/60 Hz
Power consumption
< 400 VA
Protective system - basic device
IP20
Protective system - control panel
IP40
Dimensions (L x W x H)
535 x 350 x 339 mm
Weight
30 kg
Noise level dB(A)
< 70
Noise level dB(A)
< 56
Noise level dB (A) 0.5 m distance
< 56
Audio alarm dB(A)
90
Contamination level (to IEC 60664-1)
2
Overvoltage category (to IEC 60664-1)
II
Mains connection line
3 m
Max. inlet pressure (Modul1000)
Max. inlet pressure (Modul1000b)
0.4 mbar
3.0 mbar
Minimum detectable Helium leak rates
in vacuum mode (ULTRA)
< 5
×
10
-12
mbar
l/s
Minimum detection limit in sniff mode
< 5
×
10
-8
mbar
l/s
Maximum allowed gas flow in the sniffer line
25 sccm
Maximum Helium leak rate which can be displayed
0.1
mbar l/s
Measurement range
12 decades
Maximum pumping speed (helium) at inlet
ULTRA mode
2.5 l/s
Time constant of leak rate signal (63% of the final value) < 1 s
Detectable masses
2, 3 and 4
Run-up time (after starting)
d
3
min.
Mass spectrometer
180° magn. Sector range
Ion source (2-cathodes)
Iridium/Yttrium oxide
Inlet Port
DN25 KF
Valves
solenoid
Summary of Contents for 550-300A
Page 6: ...1 4 Content jinb80e1 h Operating Manual 1501 ...
Page 10: ...1 8 Operating instructions jinb80e1 h Operating Manual 1501 ...
Page 18: ...2 8 Important safety instructions jinb80e1 h Operating Manual 1501 ...
Page 26: ...3 8 Description of equipment jinb80e1 h Operating Manual 1501 ...
Page 111: ...Technical Data 9 23 jinb80e1 h Operating Manual 1501 9 7 Commander mode Fig 9 14 Flow Chart ...
Page 112: ...9 24 Technical Data jinb80e1 h Operating Manual 1501 9 8 CE Certificate Fig 9 15 ...
Page 122: ...10 10 Error Messages and Warnings jinb80e1 h Operating Manual 1501 ...
Page 124: ...11 2 Ordering Information jinb80e1 h Operating Manual 1501 ...