Sniff mode tips
The Modul1000 Sniffer version can be used as vacuum leak detector or as sniffer
leak detector.
To use it as sniffer leak detector, the optimized sniffer line SL200 has to be
connected to „Sniff“ on the right side of the
Modul1000. In Measuring mode, the
module draws a constant gas flow through the sniffer line. The helium in the gas flow
is indicated as leak rate.
In Sniff mode, the detection limit is restricted to 1
×
10
-7
mbarl/s due to the high atmos-
pheric helium background.
The gas flow rate though the sniffer line is approximately 25 sccm.
It is electrically connected by the "ACCESSORIES“ connector.
The operating mode must be set to "Sniff mode".
In "SNIFF"-mode, the red LED on the sniffer handle indicates a bad test sample and
the green LED indicates a good test sample.
The active push-button on the sniffer handle of the sniffer line can be used for switch-
ing on the external background suppression. If the push-button is pressed for a
longer time (3 seconds), the external background suppression is switched off again.
Summary of Contents for 550-300A
Page 6: ...1 4 Content jinb80e1 h Operating Manual 1501 ...
Page 10: ...1 8 Operating instructions jinb80e1 h Operating Manual 1501 ...
Page 18: ...2 8 Important safety instructions jinb80e1 h Operating Manual 1501 ...
Page 26: ...3 8 Description of equipment jinb80e1 h Operating Manual 1501 ...
Page 111: ...Technical Data 9 23 jinb80e1 h Operating Manual 1501 9 7 Commander mode Fig 9 14 Flow Chart ...
Page 112: ...9 24 Technical Data jinb80e1 h Operating Manual 1501 9 8 CE Certificate Fig 9 15 ...
Page 122: ...10 10 Error Messages and Warnings jinb80e1 h Operating Manual 1501 ...
Page 124: ...11 2 Ordering Information jinb80e1 h Operating Manual 1501 ...