Operating and Maintenance Instruction FCU 8000-1 series
Doc.: 3109776 Ed.: 01/13.02.2002
HYDAC Filtertechnik GmbH
E
Page 106
7.3.3 Start measurement with averaging interval
Description
In this mode of operation, by contrast to point 7.3.1 "Inputting the measurement point
designation", an averaging interval can also be pre-set. Any value between 1 min and
1440 min (24 hours) in increments of 1 min can be programmed as the averaging interval.
Each current individual measurement value is always displayed for information on the
FCU display, however, only the average values of the measurements taken within the pre-
set interval are
stored
and
printed out
.
Operation
MEMORY key
Measurement point
Averaging interval
Selective deletion
Delete all
ok
start
Change meas. point
Typical applications
Continuous long term monitoring and logging of the oil cleanliness on hydraulic systems.
memory
MEMORY
Meas. point
MEMORY
Averaging interval
ok
start
Averaging interval:
10
min
+
-
0
1 2 3
4 5 6
7 8 9
or
stop
ok
start
MEMORY
Averaging interval
ok
start