Operating and Maintenance Instruction FCU 8000-1 series
Doc.: 3109776 Ed.: 01/13.02.2002
HYDAC Filtertechnik GmbH
E
Page 112
Pump pre-run time
The user has the possibility with this function, when
measuring
- with the aid of Conditioning Unit KE 1310 or
- with the aid of an external filtration unit (e.g. OF5C),
to run the pump for a limited time, during which a flow
through the FCU must start.
If the FCU detects no flow after the pump pre-run time has
elapsed, measurement is stopped and the unit to which it is
connected is switched off to prevent damage due to dry
running.
Measurement restarts as soon as enough flow is available.
The stopping time ranges from 1 to 200 sec.
Recommendation: 60 sec
Date / Time
Date and time are displayed and can be altered.
Bus address
Standard setting is 1, this setting should not be altered.
If several units with DIN measurement bus interfaces
(model code / - BUS) are connected to one bus, a bus
address between 1 and 31 must be allocated to each unit.
The same address must not be allocated twice.
Operating hours
The operating hours of the unit are displayed.
Battery voltage
The actual charging condition of the batteries is displayed.
BS pressure
supply
Selection of the compressed air supply of the optional
Bottle Sampling Unit BSU (compressed air ring main or
compressor).
Additional test
yes
In Mode M6: Bottle Sampler, a further optional individual
measurement can be started once the pre-set number of
individual measurements has been completed.
no
In Mode M6: Bottle Sampler, immediately after the pre-set
number of individual measurements has been completed,
the pressure chamber is vented and the measurement tank
drained. No further individual measurement is possible.