Operating and Maintenance Instruction FCU 8000-1 series
Doc.: 3109776 Ed.: 01/13.02.2002
HYDAC Filtertechnik GmbH
E
Page 109
7.4.1 Printing online logs during active measurement
Description
Documentation of continuous measurement by printing out the results via the built-in
printer.
Typical application
The flushing process on a hydraulic unit is to be documented over the whole cleaning
period, until a specific cleanliness class is achieved.
Operation
Logs
ok
start
Print-out format
Table / graph
ok
start
Print-out of
Particle / NAS / ISO
+
-
ok
start
+
-
Online print-out
Online print-out:
off/on
+
-
ok
start
7.4.2 Printing-out stored logs
Description
Using this mode of operation enables the user to call up measurements carried out in the
past and to print them out. The sub-menu enables selection according to different search
functions.
·
The log number (is stored continuously as a consecutive number)
·
The date (is stored automatically at the same time as the measurement)
·
The measuring position (must be input by the user)
·
By date and measuring position
Typical application
To compare the actual condition data relating to oil cleanliness in a hydraulic system,
previous measurement data must be accessed.