Operating and Maintenance Instruction FCU 8000-1 series
Doc.: 3109776 Ed.: 01/13.02.2002
HYDAC Filtertechnik GmbH
E
Page 97
7.2 Bottle sampling operation
·
Connect the FCU to the BottleSampling Unit BSU 8110:
·
Remove the filter insert (please see Point 9.1 Filter element)
·
Fit the bottle sampling adaptor in the FCU.
·
Connect the hydraulic ports (these are physically different, therefore they cannot be
connected incorrectly) to the BSU and shut off the "Outlet" port on the FCU with a
bayonet quick release plug.
·
Insert the control plug of the BSU into the matching socket of the FCU.
·
Insert the jack plug in the voltage supply socket of the FCU.
(A more detailed description can be found in the operating instructions for the BSU 8110)
7.2.1 Diagram of the BottleSampling Unit BSU 8110
V1
FCU 8110
pressure chamber
suction tube
measurement tank
M
V3
S
V2
R
sample bottle
drip tray
P
Figure 5 : Diagram of the BSU
P
: Compressed air inlet (max. 6 bar / 90 psi)
V1 : Valve for pressurizing / depressurizing the pressure chamber
V2 : Shut-off valve
V3 : Drain valve
R
: Control valve (control knob on front of BSU)
M : Pressure gauge (on front of BSU)
S
: Flow rate sensor