94
Digitizer Command Reference
TRIGger
The TRIGger command subsystem controls the behavior of the trigger system.
Subsystem Syntax
TRIGger
[:IMMediate]
:LEVel<channel>
<level> | MIN | MAX
:LEVel<channel>?
:MODE
NORMal | MASTer0,2,4,6 | SLAVe0,2,4,6
:MODE?
:SLOPe[<n>]
POS | 1 | NEG | 0
:SLOPe[<n>]?
:SOURce[<n>]
OFF | BUS | EXT | HOLD | IMMediate |
INTernal1-4 | TTLT0-7
:SOURce[<n>]?
[:IMMediate]
TRIGger[:IMMediate] causes the instrument to transition to the wait-for-sample
state immediately, regardless of the trigger source selected. The instrument must be
initiated (INITiate command) and be in the wait-for-trigger state when TRIG:IMM
is executed. A “Trigger ignored” error will be generated if the instrument has not
been initiated prior to this command or if it is not in the wait-for-trigger state.
Comments
•
Executable when initiated: Yes
•
Coupled command: No
•
Reset (*RST) condition: none
:LEVel
TRIGger:LEVel<channel>
<voltage> | MIN | MAX sets the level on the
specified channel that can be used for internally triggering the instrument.
Note
This command is valid only for TRIGger:SOURce INTernal1-4.
Parameters
Comments
•
The present range setting will determine the maximum and minimum values
that can be entered without error.
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
level
numeric
see Comments
volts