HP E1563A and E1564A Adjustments
Whatever calibration interval you select, HP recommends that complete
re-adjustment should always be performed at the calibration interval. This
will increase your confidence that the HP E1563A or HP E1564A will
remain in specification for the next calibration interval. This criteria for
readjustment provides the best measure of the digitizer’s long-term stability.
Performance data measured this way can easily be used to extend future
calibration intervals.
Aborting
an Adjustment
in Progress
Sometimes it becomes necessary to abort an adjustment once the procedure
has been initiated. Issuing a remote interface device clear command will
abort the adjustment in progress.
Caution
Never turn off mainframe power while the digitizer is making an
adjustment. If power is removed during a zero adjustment,
ALL calibration memory may be lost. If power is removed
during any gain adjustment, calibration memory for the
function being adjusted may be lost.
Adjustment Procedures
Adjustment
Conditions
See Table D-1, Recommended Test Equipment, for test equipment
requirements. For optimum performance, all adjustment procedures should
comply with following test conditions:
•
Ambient temperature of the test area is between 18
°
C and 28
°
C and
stable to within ±1
o
C.
•
Ambient relative humidity of the test area is <80%.
•
Must have a one hour warm-up with all input signals removed.
•
Shielded twisted Teflon
insulated cable or other high impedance, low
dielectric absorption cable is recommended for all measurements.
•
Keep cables as short as possible. Long test leads can act as an antenna
causing pick-up of ac signals and contributing to measurement errors.
General Procedure
Follow each adjustment by a performance verification test for added
confidence. We recommend the following general procedure.
•
Perform the Zero Adjustment Procedure.
•
Perform the Gain Adjustment Procedure(s).
•
Perform the Performance Verification Tests.