142
HP E1563A and E1564A Digitizer Error
-300
Device-specific error
-311
Memory error
-312
PUD memory lost
-313
Calibration memory lost
-330
Self-test failed
The digitizer’s complete self-test failed from the remote interface (
*TST?
command). In addition to this error, more specific self-test errors are also
reported. See also “Self-Test Errors,” following this section.
-350
Too many errors
The error queue is full because more than 20 errors have occurred.
No additional errors are stored until you remove errors from the queue.
The error queue is cleared when power has been off, or after a
*CLS
(clear status) command has been executed.
-410
Query
INTERRUPTED
A command was received which sends data to the output buffer, but the
output buffer contained data from a previous command (the previous data is
not overwritten). The output buffer is cleared when power has been off, or
after a
*RST
(reset) command has been executed.
-420
Query
UNTERMINATED
The digitizer was addressed to talk (i.e., to send data over the interface) but
a command has not been received which sends data to the output buffer. For
example, you may have executed a
SAMPle:COUNt <
count>
command
(which does not generate data) and then attempted an
ENTER
statement to
read data from the remote interface.
-430
Query
DEADLOCKED
A command was received which generates too much data to fit in the output
buffer and the input buffer is also full. Command execution continues but all
data is lost.
-440
Query
UNTERMINATED
after indefinite response
The
*IDN?
command must be the last query command within a command
string. Example:
*IDN?;:SYST:VERS?
300
Not yet implemented