HP E1563A and E1564A Digitizer Error Messages
141
-170 to -178
Expression errors
The digitizer does not accept mathematical expressions.
-211
Trigger ignored
A Group Execute Trigger (
GET
) or
*TRG
was received but the trigger was
ignored. Make sure the digitizer is in the “wait-for-trigger” state before
issuing a trigger, and make sure the correct trigger source is selected.
-213
Init ignored
An
INITiate
command was received but could not be executed because
a measurement was already in progress. Send a device clear to halt a
measurement in progress and place the digitizer in the “idle” state.
-214
Sample Trigger deadlock
You .....
-215
Arm Trigger deadlock
You .....
-221
Settings conflict
You tried to set a pretrigger count that exceeds the sample count -1. Or you
enabled one of the internal triggers as the source for a particular channel
such as channel 2 (
TRIG:SOUR INT2
) and then tried to enable one of the
limit checking features on channel 2 (
CALC2:LIM:UPP:STAT ON
).
-222
Data out of range
A numeric parameter value is outside the valid range for the command.
Example: digitizer is on the 1V range and you send
TRIG:LEV -3
-224
Illegal parameter value
A discrete parameter was received which was not a valid choice for the
command. You may have used an invalid parameter choice.
Examples:
CAL:SOUR TTLT2
(
TTLT2
is not a valid choice) or
SAMP:COUN ON
(
ON
is not a valid choice).
-230
Data corrupt or stale
-231
Data questionable
-240
Hardware error
-241
Hardware missing