Digitizer Application Information
31
SLAVe Mode
Use
TRIGger:MODE
SLAVe0
to configure a module as a slave to a
MASTer0 module. MASTer0 and SLAVe0 modules share TTL trigger lines
TTLT0 and TTLT1. TTLT0 carries the sample signal and TTLT1 carries
the trigger signal. The following table shows all pairs of TTL trigger lines
for each master-slave mode.
Master-Slave Diagrams
Figures 2-4 illustrates a module configured as a master module. Figure 2-5
illustrates a module configured as a slave module.
Table 2-2. Trigger Sources for Master-Slave Modes.
MASTer-SLAVe
Trigger Sources
MASTer
MODE
SLAVe
MODE
TRIG:SOUR1
TRIG:SOUR2
MASTer0
SLAVe0
TTLT1
Any source
(except TTLT0 & TTLT1)
MASTer2
SLAVe2
TTLT3
Any source
(except TTLT2 & TTLT3)
MASTer4
SLAVe4
TTLT5
Any source
(except TTLT4 & TTLT5)
MASTer6
SLAVe6
TTLT7
Any source
(except TTLT6 & TTLT7)