Digitizer Command Reference
49
scale (see CALibration:DAC:VOLTage command for the 4-Channel HP
E1564A internal calibration DAC outputs when CALibration:SOURce
INTernal is used).
•
The calibration voltage must be applied to the input connector if
CALibration:SOURce EXTernal is used. You must enter the external
calibrator voltage value with the CAL:VALue command when an external
calibration source is used. The HP E1564A 4-Channel Digitizer automatically
applies the DAC voltage to the internal calibration bus when
CALibration:SOURce INTernal is used. You must measure the DAC voltage
at the front panel pins (for CAL:SOURce INTernal) and enter that value with
the CAL:VALue command.
Parameters
Comments
•
The number of readings and sampling rate will default to 100 readings and
0.001 second sampling rate respectively, to provide averaging over an integral
number of either 50 Hz or 60 Hz power line cycles. This allows calibration to
cancel out any noise that is periodic with the power supply.
•
The 64V and 256V ranges are calibrated indirectly when the 16V range is
calibrated and the ON (1) parameter is set. If the OFF (0) parameter is active,
only the 16V range is calibrated and the 64V and 256V ranges retain their old
cal constant. This boolean ON/OFF parameter is checked and used only when
calibrating the 16V range. It is ignored when calibrating any other range.
•
All lower ranges (0.0625 through 4.0000) must be calibrated before calibrating
the 16V range and calculating new calibration constants for the 64V and 256V
ranges. The effects of the attenuators and amplifiers on the gain calibrations
for the lower ranges are extrapolated to derive a gain constant for the 64V
range and another for the 256V range.
•
The absolute maximum voltages for each range are shown in the next table.
The values are approximately 98% of full scale.
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
readings
numeric
25 to 4000 | DEFault
none
rate
numeric
1.25E-6 to reference
period * 8,388,607
| DEFault
Seconds
Voltage Range
Max DC Voltage (absolute value)
0.0625
0.061256409
0.2500
0.245025635
1.0000
0.980102539
4.0000
3.920410156