Digitizer Command Reference
47
CALibration
The CALibration subsystem allows you to calibrate your digitizer.
Subsystem Syntax
CALibration
:DAC:VOLTage <
voltage
> | MIN | MAX
:DAC:VOLTage? MIN | MAX
:DATA?
:GAIN[<
channel
>] [<
readings
> | DEF][,<
rate
> | DEF][,ON | 1 | OFF | 0]
:SOURce INTernal | EXTernal
:SOURce?
:STATe ON | 1 | OFF | 0
:STATe?
:STORe
:VALue <
voltage
>]
:VALue?
:ZERO[<
channel
>] [<
readings
>][,<
rate
>]
:ZERO[<
channel
>]:ALL? [<
readings
>][,<
rate
>]
:DAC:VOLTage
CALibration:DAC:VOLTage <voltage> | MIN | MAX is only active if the
CALibration:SOURce is set to INTernal. The voltage specified is output by the
internal DAC to the calibration bus (HP E1564A 4-Channel Digitizer only). You can
measure this voltage on the top two pins of the front panel D-sub connector (CAL-H
and CAL-L) and is used for calibrating the digitizer’s gain as the CAL:VALue.
Parameters
Comments
•
The maximum output levels are limited to the levels shown in the table below.
These are the HP E1564A DAC voltages recommended for calibrating each
range. The values are approximately 98% of full scale.
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
voltage
numeric
±0.061256409 - ±15.00
volts
Voltage
Range
Max DC Voltage
(absolute value)
0.0625
0.061256409
0.2500
0.245025635
1.0000
0.980102539
4.0000
3.920410156
16.0000
15.00
64.0000
not used
256.0000
not used
Summary of Contents for E1563A