70
Digitizer Command Reference
[SENSe:]
The SENSe command subsystem is used to change low-level parameters such as
voltage range, sweep and sweep offset points and to set the reference oscillator
source and frequency. It is also used to obtain measurement data from the module.
Subsystem Syntax
[SENSe:]
DATA? <
rdgs_per_channel
>[,
channel_list
]
DATA:ALL? <
rdgs_per_channel
>
DATA:COUNt?
DATA:CVTable?
[
channel_list
]
ROSCillator:EXTernal:FREQuency <
freq
>l
ROSCillator:EXTernal:FREQuency?
ROSCillator:SOURce INTernal | EXTernal
ROSCillator:SOURce?
SWEep:POINts
<
neg_value
> | MIN | MAX
SWEep:POINts?
MIN | MAX
SWEep:OFFSet:POINts
<
neg_value
> | MIN | MAX
SWEep:OFFSet:POINts?
MIN | MAX
VOLTage[<channel>][:DC]:RANGe
<
range
> | MIN | MAX
VOLTage[<channel>][:DC]:RANGe?
VOLTage[<channel>][:DC]:RESolution?
DATA?
[SENSe:]DATA? <rdgs_per_channel>[,channel_list] command returns voltage
formatted data from all channels (default) or only from the specified channel list.
Parameters
Comments
•
channel_list has the form (@1) or (@2), (@1,2), (@1:4) or (@1,2,3,4) or for
specific channels but not all, the format is (@1,3,4)
•
The readings are returned in an array in an interleaved configuration. That is,
the array contains the first reading from each specified channel followed by the
second reading from each specified channel. The readings are in channel
number order starting with the lowest to highest specified channel in the
channel list. For example, the channel list (@2,1) returns channel 1 readings
followed by channel 2 readings and returns the same as channel list (@1,2).
Parameter Name
Parameter
Type
Range of
Values
Default
Units
rdgs_per_channel
numeric
1 to MAX samples
which depends on size
of RAM on module
(see SAMPle:COUNt
command)
none
channel_list
numeric
1-2 (E1563A)
1-4 (E1564A)
N/A