HP E1563A and E1564A Verification Tests
149
Noise
Verification
This procedure is used to check the noise performance of the HP E1563A or HP
E1564A Digitizer. The digitizer’s internal short is applied to the H (HI) and L (LO)
input terminals of the channel being tested using the DIAG:SHORt <channel> ON
command.
1. Check the "Test Conditions" section at the beginning of this chapter.
2. Execute the command
DIAG:SHOR1 ON
to enable the internal short across the
H and L terminals of channel 1.
3. Set a sample interval of 25
µ
S by executing
SAMP:TIM 25e-6.
4. Select the first range (62 mV) shown in Table D-3.
5. Make 100 readings, sum them, divide by 100 and obtain the mean reading.
6. Calculate the standard deviation using the following formula (this is the rms
noise value). "reading
n
" represents the 100 readings where n = 1 to 100.
7. Record the rms noise value on the Performance Test Record and compare the
result to the appropriate test limit shown in the test record or the above table.
8. Repeat steps 4, 5 and 6 for each range listed in table D-3.
9. Repeat steps 3 to 7 for channel 2 on the E1563A 2-Channel Digitizer and
channels 2 through 4 on the E1564A 4-Channel Digitizer changing the
channel number in the DIAG:SHORt<channel> ON command and executing
the command prior to performing the steps.
Table D-3. Noise Verification Test Points.
INPUT
HP E1563A /
HP E1564A
Range
Error from
zero
internal
H-L short
DIAG:SHORt
command
62 mV
23
µ
V max
0.25 V
92
µ
V max
1 V
366
µ
V max
4 V
1.5 mV
16 V
5.9 mV
64 V
23.4 mV
256 V
93.8 mV
σ
Σ
reading
n
(
)
2
100 mean
(
)
2
–
99
-----------------------------------------------------------------------
=