HP E1563A and E1564A Verification Tests
155
Test
Input
Digitizer
Range
Low Limit
Measured Reading
High Limit
Meas
Uncert
Test Accuracy
Ratio
Gain Test (continued)
60V
64V
59.9796
60.0204
.00046
>10:1
-60V
64V
-60.0204
-59.9796
.00046
>10:1
100V
256V
99.9660
100.0340
.0007
>10:1
-100V
256V
-100.0340
-99.9660
.0007
>10:1
E1563A 25 kHz Filter Bandwidth Test
1V @
1 MHz
1V
no filter
-5 dB
-1 dB
N/A
N/A
1V @
25 kHz
1V
25 kHz filter
-5 dB
-1 dB
N/A
N/A
E1564A Filter Bandwidth Test (4 filters)
1V @
1 MHz
1V
no filter
-5 dB
-1 dB
N/A
N/A
1V @
1.5 kHz
1V
1.5 kHz filter
-5 dB
-1 dB
N/A
N/A
1V @
6 kHz
1V
6 kHz filter
-5 dB
-1 dB
N/A
N/A
1V @
25 kHz
1V
25 kHz filter
-5 dB
-1 dB
N/A
N/A
1V @
100 kHz
1V
100 kHz filter
-5 dB
-1 dB
N/A
N/A