148
HP E1563A and E1564A Verification Tests
Performance Verification
The procedures in this section are used to test the electrical performance of the
digitizer using the specifications in Appendix A of the HP E1563A and E1564A
User’s Manual as the performance standards.
The Performance Verification Tests are recommended as acceptance tests when the
instrument is first received. The performance verification tests should be repeated
at each calibration interval following acceptance. If the HP E1563A or HP E1564A
digitizer fails performance verification, adjustment or repair is needed.
Performance
Verification
Programs
Performance verification program source code is provided written in ANSI C. They
are located on the driver CD and are titled E1563VER.C and E1564VER.C.
Zero Offset
Verification
This procedure is used to check the zero offset performance of the
HP E1563A or HP E1564A Digitizer. The digitizer’s internal short is applied to the
H (HI) and L (LO) input terminals of the channel being tested using the
DIAG:SHORt <channel> command.
1. Check the "Test Conditions" section at the beginning of this chapter.
2. Execute the command DIAG:SHOR1 ON to enable the internal short across
the H and L terminals of channel 1.
3. Select each range in the order shown in Table D-2. Compare the
measurement results to the appropriate test limits shown in the table.
4. Repeat steps 2 and 3 for channel 2 on the E1563A 2-Channel Digitizer and
channels 2 through 4 on the E1564A 4-Channel Digitizer changing the
channel number in the DIAG:SHORt<channel> ON command.
Table D-2. Zero Offset Verification Test Points.
INPUT
HP E1563A /
HP E1564A
Range
Error from
nominal
internal
H-L short
DIAG:SHORt
command
62 mV
±
15
µ
V
0.25 V
±
61
µ
V
1 V
±
244
µ
V
4 V
±
976
µ
V
16 V
±
3.9 mV
64 V
±
15.6 mV
256 V
±
62.5 mV