:FILTer[:LPASs][:STATe]
INPut[<channel>]:FILTer[:LPASs][:STATe]
ON | 1 | OFF | 0 command is
used to enable or disable the low-pass filter on the specified channel.
Comments
Executable Command: No
Couple Command: No
Reset (*RST) Condition: filter OFF
:FILTer[:LPASs][:STATe]?
INPut[<channel>]:FILTer[:LPASs][:STATe]? command is used to query the
specified channel to determine if the low-pass filter is enabled or disabled. A return
value of “0” indicates the filter is OFF, “1” indicates the filter is ON.
[:STATe]
INPut[<channel>][:STATe]
ON | 1 | OFF | 0 is used to connect or disconnect the
input signal to the Digitizer’s measurement circuitry.
Note
OFF State Connections:
HP E1563A 2-Channel Digitizer:
INPut<channel>:STATe OFF connects the specified channel to ground.
HP E1564A 4-Channel Digitizer:
INPut<channel>:STATe OFF connects the specified channel to the internal
calibration bus (calibration DAC).
Comments
Executable when intitiated: No
Coupled command: No
Reset (*RST) Condition: all channels ON (connected)
[:STATe]?
INPut[<channel>][:STATe]? command queries the specified channel to determine
if the input signal is connected to, or disconnected from, the Digitizer’s measurement
circuitry. If connected, a “1” is returned; if disconnected, a “0” is returned.