112
HP E1563A and E1564A Digitizer
HP E1563A/E1564A Accuracy Specifications
1
Valid within the range of 0°C to 55°C. A zero offset calibration for all channels must be performed
if the instrument experiences a temperature <0°C or >55°C for these specifications to remain valid.
2
Specification is valid when tested at a temperature within ±5°C of the calibration temperature.
3
Amount of error that must be added for each °C outside of ±5°C of the calibration temperature.
Integral Non-linearity Specification
All ranges: 2.5 LSB
General Specifications
For indoor use, pollution degree 2
Operating altitude: 3000 meters or mainframe altitude specification, whichever is lower
Operating temperature: 0°C to 55°C
Relative humidity: up to 80% at 31°C, decreasing to 50% at 40°C
Range
Zero Offset
1
(with filter OFF)
Zero Offset
1
(with filter ON)
Gain
(% of reading)
Noise
(3 sigma)
Specifi-
cation
2
Temperature
Coefficient
3
Specifi-
cation
2
Temperature
Coefficient
3
Specifi-
cation
2
Temperature
Coefficient
3
Specifi-
cation
0.0625V
20
µ
V
1.9
µ
V/°C
28
µ
V
4.3
µ
V/°C
0.034%
0.0061%/°C
57
µ
V
0.25V
78
µ
V
6
µ
V/°C
110
µ
V
16
µ
V/°C
0.034%
0.0061%/°C
180
µ
V
1V
300
µ
V
15
µ
V/°C
430
µ
V
63
µ
V/°C
0.034%
0.0061%/°C
720
µ
V
4V
1.2 mV
60
µ
V/°C
1.7 mV
251
µ
V/°C
0.034%
0.0061%/°C
2.88 mV
16V
21 mV
1.3 mV/°C
21 mV
1.63 mV/°C
0.034%
0.0061%/°C
14.7 mV
64V
28 mV
1.65 mV/°C
34 mV
4.24 mV/°C
0.034%
0.0061%/°C
48 mV
256V
79 mV
4.28 mV/°C
110 mV
16.2 mV/°C
0.034%
0.0061%/°C
189 mV
Summary of Contents for E1563A