Digitizer Command Reference
105
A zero response indicates that the self-test passed. Any non-zero response indicates
that the test failed. Input the failed test number into the
TEST:ERR? <
number>
command. The returned values from this will be the result code and a string. See
Appendix B, Error Messages, for information on interpreting the result code and
string.
Comments
•
Executable when initiated: No
•
Coupled command: No
•
Reset (*RST) condition: none
*WAI
*WAI
causes the HP E1563A and E1564A Digitizers to wait for all pending
operations to complete before executing any further commands.
Comments
•
The *WAI will not wait for all measurements to complete when an INIT
command is executed to start measurements. *WAI considers INIT finished
once it is processed although the instrument may still be taking measurements.
In this case, the instrument will move on to the next command following
*WAI while measurements are being taken.
•
Executable when initiated: Yes
•
Coupled command: No
•
Related commands:
*OPC, *OPC?
•
Reset (*RST) condition: none