Digitizer Command Reference
55
DIAGnostic
The DIAGnostic subsystem contains several commands which were developed to
test the instrument at the factory. Some of these commands may prove useful for
isolating problems or for use in special applications.
Subsystem Syntax
DIAGnostic
:DAC:OFFSet[<
channel
>] <
voltage
>
:DAC:OFFSet[<
channel
>]:RAMP <
count
>
:DAC:GAIN[<
channel
>] <
value
>
:DAC:SOURce <
voltage
>
:DAC:SOURce:RAMP <
count
>
:INTerrupt:LINE 0 | 1 | 2 | 3 | 4 | 5 | 6 | 7
:INTerrupt:LINE?
:MEMory:SIZE <
size
>
:MEMory:SIZE?
:PEEK? <
reg_number
>
:POKE? <
reg_number
>,<
data
>
:SHORt[<
channel
>] ON | 1 | OFF | 0
:SHORt[<
channel
>]?
:STATus?
:DAC:OFFSet
DIAGnostic:DAC:OFFSet
[<
channel
>]
<voltage> writes the specified voltage
value to the calibration offset DAC of the specified channel when the
DAC:GAIN
command is sent. This offset voltage value is not used unless a
DAC:GAIN <
value
>
is sent to the calibration gain DAC. This command is a factory diagnostic routine.
Parameters
Comments
•
Executable when initiated: No
•
Coupled Command: No
•
Reset (*RST) Condition: None
:DAC:OFFSet:RAMP
DIAGnostic:DAC:OFFSet
[<
channel
>]
:RAMP <count> outputs to the specified
channel, a ramp of DAC values from 0 to 255 with the DAC code changing
approximately every 100
µ
S. This command is a factory diagnostic routine.
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
voltage
numeric
-2.5 to +2.5
none