Digitizer Command Reference
65
3. The instrument’s FIFO memory is filled. This can be prevented by fetching the
data from memory in blocks faster than the sample rate can fill memory.
Note
The INIT[:IMMediate] and INIT:CONTinuous commands return “1” to *OPC?
when the instrument begins measurement not when measurements complete. To
determine when a non-continuous measurement is complete, use the
DIAG:STATus? command and monitor bit 6. You can also detect when
measurements are complete by monitoring the “measurement complete” bit (bit 9)
of the STATus:OPERation:CONDition register in the STATus system. See the
STATus command. *WAI, *OPC and *OPC? will all be fulfilled immediately after
the INIT command is processed not when the measurements are complete.
Comments
•
Executable when initiated: No
•
Coupled Command: No
•
Reset (*RST) Condition: Idle state
:CONTinuous?
INITiate:CONTinuous? queries the instrument to determine if the
INITiate:CONTinuous is enabled or disabled.