
34
Digitizer Application Information
Input Overload Condition
Overload voltages may occur which will open the channel input relay
disconnecting the input signal from the channel. Overload voltage by range
is shown in the following table.
Overload Reporting
The overload is reported both when the readings are retrieved and when the
next measurement is initiated. If an overload occurred:
1. An error message is returned when data is retrieved informing you
that the data is questionable
(Overload detected - data questionable)
.
2. An error message is also returned when you initiate the next
measurement
(Overload detected - attempting re-connect of input relays).
Range
Voltage Input Condition
Vmax
62 mV to 4V
High or Low to Guard
>20V
16V to 256V
Low to Guard
>40V
Summary of Contents for E1563A
Page 8: ......
Page 24: ...24 Digitizer Module Set up ...
Page 84: ...84 Digitizer Command Reference ...
Page 110: ...110 ...
Page 138: ...138 HP E1563A and E1564A Register Based Programming Appendix B ...
Page 156: ...156 HP E1563A and E1564A Verification Tests ...